Order Name LAB318 PLATELET FUNCTION ANALYSIS

Important Note

*This test is orderable in Cerner as Platelet Function Analysis UVM

Method

PFA-100

CPT(s)

Description CPT Code
Platelet Function Analysis 85576

 

Reference Range

COL/EPI: 94 – 193 Seconds

COL/ADP: 71 – 118 Seconds

Instrumentation

PFA-100

Specimen Information

Container Specimen Temperature Collect Vol Submit Vol Min Vol Stability
Blue Top Tube Whole Blood Ambient To fill line To fill line To fill line 4 hours

Testing must be performed within 4 hours of sample collection. Requires a separate aliquot if other coagulation tests are requested.

Platelet Function Analysis (PFA) is strongly dependent on the correct method of blood collection requiring venipuncture through a19-21G needle drawn directly into a full size evacuated sodium citrate tube. Do not draw through a line or port, which may cause platelet clumping and result in specimen rejection. Ensure proper specimen mixing by gently inverting by hand 4 times. Must remain unspun at ambient temp. Testing must be performed within 4 hours of sample collection. PFA requires separate tube; additional samples must be collected if other coagulation tests are requested.

 

Result Component(s)

Reporting Name Epic Code Atlas Code Mayo Access ID LOINC 
COL/EPI Cartridge   PFAEPI   24471-5

 

Performing Location

University of Vermont Medical Center

Test Schedule / Analytical Time / Test Priority

Daily / 24 Hours / Available STAT

Section

Coagulation

Is the UVMMC lab NY State Certified to perform this testing?  Yes/No

Yes