Blood Gas Capillary
Specimen Requirements
Specimen must arrive within 30 minutes of draw.
Specimen Type: Capillary blood-Clotted blood is not acceptable.
Container/Tube: Heparinized capillary tube
Specimen Volume: 120 µL
Collection Instructions:
1. Collect capillary blood anaerobically.
2. Expel air bubbles and plug ends-Air bubble is not acceptable.
Stability
Analyze immediately after collection.
Reference Range
pH Cap: 7.35-7.45
CO2 Total Cap: reference range is not established.
pCO2 Cap: 33-47 mmHg
pO2 Cap: 80-100 mmHg
O2 Sat Cap: 95-98%
Base Excess Cap: reference range is not established.
HCO3 Cap: reference range is not established.
Instrumentation
Radiometer ABL90 Flex
Methodology
Ion-Selective Electrode (ISE)
Test Schedule/Analytical Time/ Test Priority
Daily / Same day / Available STAT
CPT (s)
82803